Event codes and recommendations
Below is a list of the event codes that your LWT instrument might display should a problem arise, and
recommended action for each event. The first letter of the code refers to NAMUR codes:
• F: Failures
• S: Function check
• C: Out of specifications
• M: Maintenance request
—
Table 4
Event codes and recommended action
No.
Event message
Recommended action
S080.000
Maximum level rate exceeded
Make sure max level rate matches the process
S081.001
GpC alarm, last valid gpc used
Check installation, GpC marker, process cond.
F203.002
Sensor has lost surface echo
Check installation and parameter adjustments.
S088.003
No interface detected
For information purposes only
S085.004
Emulsion has been detected
For information purposes only
F234.005
Electronic failure or harsh env
Check installation for vibrations or EMC interference
S060.010
pV value out of range
Check process conditions and device config.
C153.011
GpC currently in calibration
Wait for calibration to finish.
S083.012
too many echos detected
Check installation or device parameters.
F233.013
Reference echo not detected
Check installation and coupler connection.
F210.016
Invalid device parameters
Enter valid parameters
F236.017
Sensor NV Failure
the sensor should be replaced aSap
S110.018
Successful sensor reboot
Successful sensor reboot
S062.019
Surface inside blocking distance
Check process conditions and device config.
F237.020
Sensor Board communication error. Restart sensor. If condition persists, call aBB
F233.023
Electronics NV Failure
the electronics must be replaced; call aBB.
S082.024
Weak echo amplitude
Check installation and process conditions.
F232.025
Unconnected coupler or probe
Check connection of head, coupler and probe
S061.026
Surface inside safety distance
Check process conditions and/or device config.
M30.031
Maintenance required
Verify sensor functionality, clear diagn. Mess.
M011.033
temperature out of range
Check ambient temperature. Use remote option
F230.034
First coupler seal leak
Check coupler.
F235.035
Load sensor configuration failure
Restart sensor. If condition persists, call aBB
C151.036
Data Simulation warning
the Output value is derived from simulated value
C150.037
alarm Simulation warning
alarm condition is being simulated by the device
M012.042
NV replace error.
Restart sensor. If condition persists, call aBB
M035.045
hMI validation test failure
Replace the hMI; call aBB.
F231.046
Broken probe or wrong length
Replace probe or adjust probe length.
M026.047
Build-up on probe
Clean probe.
Содержание LWT300 series
Страница 1: ... USER GUIDE LWT300 series Guided wave radar level transmitter Modbus ...
Страница 36: ...Page intentionally left blank ...
Страница 40: ...Page intentionally left blank ...
Страница 66: ...Page intentionally left blank ...
Страница 70: ...Page intentionally left blank ...
Страница 76: ...Page intentionally left blank ...
Страница 96: ...Page intentionally left blank ...
Страница 98: ...Page intentionally left blank ...
Страница 104: ...Page intentionally left blank ...
Страница 105: ...Page intentionally left blank ...
Страница 106: ...Page intentionally left blank ...
Страница 107: ......