38
2
Press
Next
. The
Set PV
screen appears. This sets the type of measurement that you want to make.
a
Press
Edit
. The list of available measurements appears (default value:
Level
).
b
Scroll up or down the list and highlight the required measurement.
c
Press
OK
. The measurement is selected.
For more information, see “Setting variables” on page 45.
3
Press
Next
. The
PV Unit
screen appears. This sets the measurement unit for the measurement type
that you just previously selected.
a
Press
Edit
. The list of available measurement units appears.
b
Scroll up or down the list and highlight the required measurement unit.
c
Press
OK
. The measurement unit is selected.
For more information, see “Setting variable units” on page 45).
4
Press
Next
. The ↓
Empty/Zero
screen appears. This sets the lowest level that you expect to measure
in the vessel.
a
Press
Edit
.
b
Set the required value (see “Setting a value” on page 35).
c
Press
OK
. The value is set.
For more information, see “Changing the empty/zero value” on page 64.
5
Press
Next
. The ↑
Full/Span
screen appears. This sets the highest level that you expect to measure in
the vessel.
a
Press
Edit
.
b
Set the required value (see “Setting a value” on page 35).
c
Press
OK
. The value is set.
For more information, see “Changing the full/span value” on page 65.
6
Press
Next
. The
Max Level Rate
screen appears. This sets the expected rate at which the vessel fills
up or is drained.
a
Press
Edit
. The list of available values appears.
b
Scroll up or down the list and highlight the required value.
c
Press
OK
. The value is applied.
For more information, see “Selecting maximum level rate” on page 51.
7
Press
Next
. The
Display Line1 View1
screen appears (this sets the measurement displayed on the
first line of text, as illustrated in Figure 30 on page 40).
a
Press
Edit
. The list of available values appears (default value:
Level
)
b
Scroll up or down the list and highlight the required value.
c
Press
Next
. The value is applied and you return to the beginning of the
Easy Setup
menu.
For more information, see “Configuring the display” on page 39.
Содержание LWT300 series
Страница 1: ... USER GUIDE LWT300 series Guided wave radar level transmitter Modbus ...
Страница 36: ...Page intentionally left blank ...
Страница 40: ...Page intentionally left blank ...
Страница 66: ...Page intentionally left blank ...
Страница 70: ...Page intentionally left blank ...
Страница 76: ...Page intentionally left blank ...
Страница 96: ...Page intentionally left blank ...
Страница 98: ...Page intentionally left blank ...
Страница 104: ...Page intentionally left blank ...
Страница 105: ...Page intentionally left blank ...
Страница 106: ...Page intentionally left blank ...
Страница 107: ......