Operating S4 with an older sensor
D184B140U02 FSM4000
113
Type of
flowmeter
sensor, version
level
Meter size DN
Connection
board with
voltage splitter
(switch S903
must be closed)
FRAM design
Coil supply
current from:
Reference
voltage
SE2_, SE4
1 ... 1000
No
With flowmeter sensor
calibration values
Cs = see name plate
Cz = see name plate
Transmitter S4
automatic
DS2_
DS4_
10DS3111(A-D)
10DS3111(E)
1 ... 1000
≤
300
≤
400
≤
300
No
Cs = calculated;
Cz = 0%
Transmitter S4
from name
plate
10DI1425
≤
400
No
Cs = calculated;
Cz = 0%
Transmitter S4
90 mV
DS4_
10DS3111(E)
≥
350 ... 1000
≥
350 ... 400
No
Cs = calculated;
Cz = 0%
Transmitter S4
from name
plate
10DS3111(A-C)
10DS3111(D)
10DS3111(E)
≥
500
≥
500
≥
500
No
Cs = calculated;
Cz = 0%
External supply
power
from name
plate
10D1422,
10DI1425
3 ... 1000
≥
500
yes
yes
Cs = calculated;
Cz = 0%
External supply
power
–
10DS3111A
10D1462
10D1472
350 ... 600
150 ... 900
15 ... 100
no, use 1000
Ω
adapter board
calculated
External supply
power
from name
plate
Important
If the flow indicator shows the incorrect direction after successful startup of the meter, e.g.,
reverse instead of forward, correct this in the Operating Mode submenu of the transmitter.
First switch off the programming protection ("Prog. Level"
→
"Specialist"). Then select the
parameter "Directional display" under the parameter "Operating Mode", and change "normal"
to "inverse". Finally, reactivate programming protection via "Prog. Level"
→
"Locked").