
Parameterization
D184B140U02 FSM4000
97
7.4.4 Readjusting the coil temperature
If a readjustment is required, make sure that the flowmeter sensor and coil are relatively close
to ambient temperature. You might need to allow the flowmeter to cool off overnight.
Work steps:
• Switch on the transmitter.
• Under "Diagnostics / Adjustment / CabLength", enter or check the cable length.
• Enter the ambient temperature in the submenu "Diagnostics / Adjustment / Coil Temp".
• Store the new reference resistance in the submenu "Diagnostics / Adjustment / Transfer
RT20".