Messages
and
tests
D184B140U02 FSM4000
103
8.3
Warning messages during operation
Warning code and
ID letter
Prior-
ity
Description
Possible cause
Corrective action
Warning: 1
Q Simulation
2
Before operation with flow, a
flowrate can be simulated. The
output values correspond to the
simulated flowrate entered.
Simulation mode on.
After completing the
simulation program,
switch off the
parameter "Simulation
Mode".
Warning: 2
Totalizer reset
1
All totalizers (forward/reverse,
diff. and overflow counters are
reset.)
External totalizer
reset was performed.
Open the switch at the
contact input (terminal
81, 82).
Warning: 3
Test Mode
3
Only for checking the transmitter
using Simulator 55XC4000.
Test Mode on.
After running the
simulation on the
simulator, switch off the
parameter "Test Mode".
Warning: 4
Function test
4
If during HART communication,
the functional test for the contact
output or contact input was
started, the Warning 4 is
displayed.
Function test on.
After completing the
self check test, exit the
routine.
Warning: 7
1)
Ext.Dat.loaded
9
Transmitter has recognized
different flowmeter sensor data
and loaded it in the internal
FRAM. Loaded are the system
and flowmeter sensor data.
Defective FRAM,
repair unit, replace
instrument.
Write down all totalizer
values, totalizers
should be reset.
Warning: 8a
1)
Update int.Dat
10
Transmitter has detected an error
in the internal FRAM and has
repaired the data with content of
the external FRAM.
Defective FRAM,
repair unit, replace
instrument.
Check the setup data
and correct, if
necessary.
Warning: 8b
1)
Update ext.Dat.
Transmitter has detected an error
in the external FRAM and has
repaired the data with content of
the internal FRAM.
Defective FRAM or
data was modified.
Check the setup data
and correct, if
necessary.
Warning: 9a
Overflow > F
5
Forward totalizer has reached
the maximum count and has
restarted at zero.
Max. totalizer
overshot, 1 overflow
as value was
totalized.
Reset totalizer, if
necessary.
Warning: 9b
Overflow < R
6
Reverse totalizer has reached
the maximum count and has
restarted at zero.
Max. totalizer
overshot, 1 overflow
as value was
totalized.
Reset totalizer, if
necessary.
Warning: 9c
Overflow Diff.
7
Diff. totalizer has reached the
maximum count and has
restarted at zero.
Max. totalizer
overshot, 1 overflow
as value was
totalized.
Reset totalizer, if
necessary.