364xS User’s Manual
44
Operation
Display indication
Description
Level
Type
Calibrate
This function provides all the Calibration settings
Main Menu
Sensor Trim
This function provides the trimming of the sensor
Sub Menu
Lower sensor trim
Set the lower Sensor Trim point
Utility
Numerical
Upper sensor trim
Set the upper Sensor Trim point
Utility
Numerical
Output Trim
This function provides all the trimming of the
current output
Sub Menu
Set 4 mA
Trim the output current to 4 mA when the action is
performed
Utility
Numerical
Set 20 mA
Trim the output current to 20 mA when the action is
performed
Utility
Numerical
Reset to Factory Trim
Set the output current Trim and the Sensor Trim at
the Calibration performed in the factory during the
trasmitter manufacturing
Utility
Enumerated
Diagnostics
This function provides all the Diagnostics
information
Main Menu
Diagnostics
This function provides all the Main Diagnostics
information
Sub Menu
PV
Monitor the PV value
Utility
Numerical
HART Eng. Output
Monitor the HART Engineering Output value
Utility
Numerical
Output current
Monitor the Current Output value
Utility
Numerical
Output %
Monitor the Percentage Output value
Utility
Numerical
Permanent Totalizer
Monitor the Permanent Totalizer value
Utility
Numerical
Batch Totalizer
Monitor the Batch Totalizer value
Utility
Numerical
Static Pressure
Monitor the Static Pressure value
Utility
Numerical
Sensor temperature
Monitor the Sensor temperature value
Utility
Numerical
Loop Test
This function provides to force the output current at
a specific value
Sub Menu
Set 4 mA
Set the output to 4 mA
Utility
Auto Adjust
Set 20 mA
Set the output current to 20 mA
Utility
Auto Adjust
Set custom value
Set the output current to a defined value
Utility
Manual Adjust
Digital LCD Integral Display view setting
To view the menu in a specific language:
Navigate the Digital LCD Integral Display menu, select
Display
>
Language
, and the used language will be
available. To modify select
Edit
then scroll the desired language with the up and down keys and confirm with
OK
.
)
Digital LCD Integral Display Keys sequence to set the language: 3 – 1 – Edit
To modify the Operator view:
Navigate the Digital LCD Integral Display menu, select
Display
>
Operator View
, and the used view will be
available. To modify select
Edit
then scroll the desired Operator view (see the following table) with the up
and down keys and confirm with
OK
.
One Line
to view the Variable 1 as defined in the Display
One Line + Bargraph
to view the Variable 1 as defined in the Display and the Bargraph
proportional with the variable full scale
Two Lines
to view the Variable 1 and the Variable 2 as defined in the Display
Two Lines + Bargraph
to view the Variable 1 and the Variable 2 as defined in the Display
and the Bargraph proportional with the variable full scale
Содержание 2600T Pressure Transmitter Series
Страница 1: ...InstrumentationIT 2600T Pressure Transmitter family ABB User s Manual 364DS and 364PS Models ...
Страница 7: ...364xS User s Manual 7 This page has been intentionally left blank ...
Страница 72: ...364xS User s Manual 72 Additional Safety ...
Страница 74: ...364xS User s Manual 74 Additional Safety ...
Страница 75: ...364xS User s Manual 75 Digital LCD Display Keys sequences Annex I Digital LCD Integral Display Menu Flow Chart ...