AD4329A-DLC
Page 27
7. Basic Weighing Function
7.1. Weighing mode
When the AD4329A-DLC is turned on, all the segment of the display illuminate/extinguish for checking the display.
Then the AD4329A-DLC is in weighing mode and starts weighing.
The contents in the display in weighing mode are as follows.
・
Weighing value
・
Positive overload
・
Negative overload
Error display
Description
errAd n
The n-th digital load cell has a problem in the internal circuit.
errlC n
The n-th digital load cell doesn't respond.
Check if serial number setting is correct.
errov n
The load on the n-th digital load cell exceeds the rated capacity.
Check the actual load.
-------
The load exceeds the power-on zero range.
Press [MODE] to check the current load.
.
-
.
1234.56