AD4329A-DLC
Page 44
12.1. Confirming the digital load cell
The display shows the number of load cells connected in the
box indicated by a dotted line.
Press [TARE] to select the load cell number to be confirmed.
The example at the right is when the 4th load cell is selected.
After displaying "
lC4 5t
" for 2 seconds, its status is
displayed.
When an error occurs, the corresponding "
o
" on the display
is switched to "
1
".
After displaying "
lC4 5n
" for 2 seconds, the serial number is
displayed.
After displaying "
lC4 Cap
" for 2 seconds, the maximum capacity
is displayed.
After displaying "
lC4load
" for 2 seconds, the actual load applied
to the load cell is displayed.
After displaying "
lC4 4C
" for 2 seconds, the value calibrated is
displayed.
After displaying "
lC4 temp
" for 2 seconds, the internal
temperature (
℃
) is displayed.
AD error
Positive overload
Negative overload
Out of positive operating
(allowable) temperature range
Out of negative operating
(allowable) temperature range
No response
1ooooo
CHcl1-4
lC"1
oooooo
2900414
26.2
0
200
Change
lC"4
10000
CHc
1
lC