AD4329A-DLC
Page 41
11.1. Output data format
When F-41 is 0 / 1 / 2, the output data format is as follows.
●
F-45 = 0: format 1
Item
Value
ST, stable
US, unstable
Header 1
OL,
The weighing data is
overload
G<SP>, gross
value
N<SP>, net
value
T<SP>, tare
value
Header 2
PT,
The weighing data is
preset tare value
+01234.5
Numerical value with polarity
+00000.0 0
+<SP><SP><SP><SP>.<SP><SP> Positive
overload
Weighing data *1
-<SP><SP><SP><SP>.<SP><SP> Negative
overload
kg Kilograms
Unit
(
CF-001
)
*1
<SP>t Ton
*1) it is possible to select the number of digits and decimal point format for output data (CF-010).
0: Two digits, "." dot
1: Three digits, "." dot
2: Two digits, "," comma
3: Three digits, "," comma
●
F-45 = 1: format 2
If overload appears on the display, +99999999 without the decimal point will be output.
Weighing data
Terminator
+ 0 0 1 2 3 4 . 5
<CR> <LF>
S T , G
<SP>
, + 0 1 2 3 4 . 5 k g
<CR>
<LF>
Header 1
Weighing data
Header 2
Unit
Terminator