16.3. Comparator
Function No.
Setting range
Description
Default value User setting
f1-01
-999999
to 999999
Upper limit value of comparator.
Decimal point position depends on
Cf-01
.
0
f1-02
-999999
to 999999
Lower limit value of comparator.
Decimal point position depends on
Cf-01
0
f1-03
0 to 2
Comparator mode:
0
: No comparison
1
: Comparison, excluding the zero band
2
: Comparison, including the zero band
2
f1-04
-999999
to 999999
Set the zero band for the comparator mode.
0
f1-05
1 to 3
Hysteresis direction:
1
: Upward 2-level judgment
2
: Upper / lower limit judgment
3
: Downward 2-level judgment
2
f1-06
0.0 to 5.0
Hysteresis time. (Unit: 0.1 second)
When
0.0
, the hysteresis mode is not used.
0.0
f1-07
00 to 99
Hysteresis width. (Unit: d)
When
00
, the hysteresis mode is not used.
99
16.4. Analog output
Function No.
Setting range
Description
Default value User setting
f2-01
-999999
to 999999
Measured value at DAV 0V output.
Decimal point position depends on
Cf-01
.
0
f2-02
-999999
to 999999
Measured value at DAV 10V output.
Decimal point position depends on
Cf-01
.
10000
f2-03
-999999
to 999999
Measured value at DAI 4 mA output.
Decimal point position depends on
Cf-01
.
0
Ff2-04
-999999
to 999999
Measured value at DAI 20 mA output.
Decimal point position depends on
Cf-01
.
10000