9.4. Power on zero
Select the zero adjustment operation when the power is connected. (
Cf-10
)
0:
Digital zero function off
Measurement is based on the zero point of the calibration.
Use for measuring an absolute value such as force measurement.
1:
Perform digital zero
Measurement is based on the zero point when the power is connected.
Use for weighing a load.
2:
Use the state when the power is disconnected.
Measurement is based on the DZ.
Use for measurements that require a long time such as a hopper scale.
It is a useful function in case of a power failure.
9.5. Backup of DZ
The DZ value is saved in non-volatile memory (EEPROM).
Writing up to 10,000,000 times is available, so keep track of use frequency.
If the available DZ memory is 10% or less, the
DZ
LED blinks.
With the setting (
Cf-10
=
2
), the backup memory is updated only when the DZ value is
updated by key operation or command.
(The zero point updated using the zero tracking function will not be saved in non-volatile
memory.
DZ
on
Power
supply
on
Power
supply
off
Change of power
supply
Change of input
0: Digital zero
function off
1: Perform digital
zero
2: Use state when
the power is
disconnected
0
0
0
0
ON
OFF
Change of display
Difference by
the setting