23
7.1.2. Span calibration mode
Input the value to be displayed when the actual load for span
calibration is applied to the load cell.
Wait for the
HOLD
LED to turn on and press the
key.
Select the digit to be changed.
Increase the value of the digit to be changed.
Change the polarity.
Perform span calibration and proceed to the storing mode.
Cancel span calibration and proceed to the storing mode.
* After span calibration, the AD-4531B displays the mV/V value of span calibration for
3 seconds, and then proceeds to the storing mode.
7.1.3. Storing mode
Save the calibration zero, span and displayed value acquired.
When calibration is not performed, data is not saved.
Save the data acquired and return to the measurement mode.
Do not save the data acquired and return to the
measurement mode.
Measurement
mode
Storing mode