16.2. Basic functions
Function No.
Setting range
Description
Default value User setting
f0-01
00000
to 11111
Disable key
Each digit of the setting corresponds to a key
switch. Only available in the measurement mode.
Key assignment:
0
: Enabled
1
: Disabled
0 0 0 0 0
00000
(Binary)
f0-02
0 to 9
Digital filter. Cutoff frequency:
0
: Off
5
: 2.8 Hz
1
:11 Hz
6
: 2 Hz
2
: 8 Hz
7
: 1.4 Hz
3
: 5.6 Hz
8
: 1 Hz
4
: 4 Hz
9
: 0.7 Hz
8
f0-03
1 to 20
Display update rate:
1
: 1 time/second
10
: 10 times/second
2
: 2 times/second
20
: 20 times/second
5
: 5 times/second
20
f0-04
0 to 4
Hold mode:
0
: Off
3
: Bottom hold
1
: Sample hold
4
: Bipolar peak hold
2
: Peak hold
1
f0-05
0.0 to 9.9
Hold averaging time:
Set by the unit of 0.1 second.
When
0.0
, averaging is not performed.
0.0
f0-06
0000
to 1111
LATCH function
Function corresponds to an external input latch.
Setting and latch
Assignment:
0
: Off
0 0 0 0
1
: On
Displayed value latch
Comparator latch
Analog output latch
Serial output latch
0000
(Binary)
f0-07
0 to 6
External
input 1
0
: Off
4
: Start HOLD
1
: ZERO
5
: Stop HOLD
2
: HOLD
6
: LATCH
3
1
f0-08
0 to 6
External
input 2
2
f0-09
0 to 9
External
output 1
0
: Off
5
: OK
1
: DZ
6
: LO
2
: During HOLD
7
: Measuring (ON)
3
: HOLD busy
8
: Measuring (1 Hz)
4
: HI
9
: Measuring (50 Hz)
1
f0-10
0 to 9
External
output 2
2