7.2. Calibration errors
Display
Cause
Remedy
C?e2
Voltage at zero point calibration
exceeds in the positive direction.
Confirm the rating and connection of
the load cell.
C?e3
Voltage at zero point calibration
exceeds in the negative direction.
C?e4
The value of the calibration weight
exceeds the rated capacity.
Use a proper calibration weight.
C?e5
The value of the calibration weight
is less than the minimum division.
C?e6
The load cell sensitivity is
insufficient.
Confirm the load cell connection.
Use a proper calibration weight.
C?e7
Voltage at span calibration is less
than voltage at the zero point.
Confirm the load cell connection.
C?e8
The load cell output voltage is too
high when loaded to capacity.
Use a load cell with a greater rated
capacity or set a smaller rated
capacity value.