22
7. CALIBRATION
The AD-4531B measures voltage signals from sensors and displays the values.
Calibration is performed so that the AD-4531B performs correctly.
The decimal point (
Cf-01
), minimum division (
Cf-02
) and rated capacity (
Cf-03
) are
set using the function mode.
The zero point input voltage (
Cf-04
), the span input voltage (
Cf-05
) and the displayed
value for the span input voltage (
Cf-06
) are adjusted using the calibration mode.
Calibration setting by the function mode is also available. (Digital span)
* During calibration, maintain a stable environment to prevent calibration errors.
* When the measured value is stable, the
HOLD
LED is turned on.
* The decimal point blinks to indicate that the current value is not a measured value.
7.1. Calibration modes
In the measurement mode, press the
+
keys to enter the calibration mode.
Enter the zero point calibration mode.
Return to the measurement mode.
7.1.1. Zero point calibration mode
With nothing on the load cell, wait for the
HOLD
LED to turn
on and press the
key.
Perform zero point calibration and
proceed to the span calibration mode.
Cancel zero point calibration and
proceed to the span calibration mode.
Hold down to display the mV/V value of the zero point.
Span calibration