* Each of the A and B connections has two terminals.
These terminals are connected internally and can be used for a terminating resistor or
multi-drop connection.
5.1.6 Analog output connector (Option)
(1) COM
Common terminal of the analog output
(2) IOUT
Analog current output terminal
(3) VOUT
Analog voltage output terminal
5.2. Equivalent circuit diagram of the control input section
When connecting the input terminal:
Select a switching element of the non-voltage input type, such as a
mechanical contact or semiconductor contact.
Use the leakage of the switching element, when OFF, of 30
μA or less
Toggle switch
Mechanical relay
contact
Semiconductor
contact
(Transistor open
collector)
TTL open collector output
+5V
3.3k
Ω
200
Ω
AD-4531B inner side
0.1
μ
F
Machine contact
(Switch)
Keep each ON and OFF time
for 100 ms or more
.
Approx. 1.5mA
(1) EXT IN 1
(3) IN COM
* The circuit of (2) EXT IN 2 is the same as the
circuit of (1) EXT IN 1.