56
TABLE 6.2 J2 INPUT / OUTPUT JUMPER CONFIGURATION OPTIONS
FOR GFM SERIES FLOW METERS
PV INPUT
TYPE
(TIO INPUT)
J2 JUMPER CONFIGURATION
GFM
CABLE KIT
NOTE
J2A
J2B
J2C
J2D
J2E
J2F
0 - 5 Vdc
2 - 3
5 - 6
8 - 9 10-11 14-15 17-18
KIT-TM-DD
KIT-TM-FD
5 - 10 Vdc
2 - 3
5 - 6
8 - 9 11-12 14-15 17-18 KIT-TM-RD
+5Vdc
reference
signal must
be used
0 - 10 Vdc
2 - 3
5 - 6
8 - 9 11-12 14-15 17-18
NOT
SUPPORTED
FOR GFM
Special Order
option! (PCB
hardware must
be changed)
4 – 20 mA
2 - 3
5 - 6
8 - 9 10-11 13-14 16-17
N/A (user
custom cable
assembly)
(249 Ohm
passive, not
isolated current
input)
d) Parameters Configuration
Following parameters must be configured:
•
Device Function
(See 4.3.12 Submenu “Device Function”). “Meter”
function must be selected.
•
Full Scale Range
(See 4.3.14 Submenu “Device Calibration”). Full
Scale Range parameter must be set equal to the GFM full scale flow
rate in Litr/min.
•
Fluid Std. Density
(See 4.3.14 Submenu “Device Calibration”). This
parameter is required only when mass-based engineering units are
selected.
User may configure other parameters (see Paragraph 5.3) according
to individual preferences and application requirements.
NOTE: If “Full Scale Range”, “Device Function” and “Fluid Std.
Density” parameters are not set properly the device may have
erroneous reading and unpredictable behavior.
,
Содержание TIO
Страница 14: ...10 Figure 4 3 RS 485 Multidrop Half Duplex Two Wire System...
Страница 55: ...51 Figure 6 2 Connecting TIO to the GFM using 0 5 Vdc output from DB9 connector...
Страница 56: ...52 Figure 6 3 Connecting TIO to the GFM using 5 10 Vdc output from RJ11 connector...
Страница 57: ...53 Figure 6 4 Connecting TIO to the GFM using 4 20 mA output from DB9 connector...
Страница 61: ...58 Figure 6 6 Connecting TIO to the GFC Using 0 5Vdc Input Output from DB15 Connector...
Страница 62: ...59 Figure 6 7 Connecting TIO to the GFC Using 4 20mA Input Output from DB15 Connector...
Страница 68: ...65 Figure 6 9 Connecting TIO to the Generic Flow Meter...
Страница 69: ...66 Figure 6 10 Connecting TIO to the Generic Flow Controller...
Страница 95: ...92 APPENDIX D Mechanical Drawings...
Страница 96: ...93 APPENDIX E Circuit Layout Diagrams Circuit Layout Top...
Страница 97: ...94 Circuit Layout Bottom Mirror...