RS-232 Communication Interface Connection:
Crossover connection must be established:
RS-232 RX
(pin 2 on the host PC DB9 connector)------pin 7 of the 9 pin "D" connector (TX-)
RS-232 TX
(pin 3 on the host PC DB9 connector)------pin 3 of the 9 pin "D" connector (RX+)
RS-232 SIGNAL GND
(pin 5 on the host PC DB9 connector)------pin 6 of the 9 pin "D" connector
RS-485 Communication Interface Connection:
The RS485 converter/adapter must be configured for: multidrop, 2 wire, half
duplex mode (See Figure 4.3). The transmitter circuit must be enabled by TD or
RTS (depending on which is available on the converter/adapter). Settings for the
receiver circuit should follow the selection made for the transmitter circuit in order
to eliminate echo.
RS-485 T(-) or R(-)
----------------- pin 7 of the 9 pin "D" connector (TX-)
RS-485 T(+) or R(+)
----------------- pin 3 of the 9 pin "D" connector (RX+)
RS-485 GND (if available) ----------------- pin 6 of the 9 pin "D" connector
Содержание TIO
Страница 14: ...10 Figure 4 3 RS 485 Multidrop Half Duplex Two Wire System...
Страница 55: ...51 Figure 6 2 Connecting TIO to the GFM using 0 5 Vdc output from DB9 connector...
Страница 56: ...52 Figure 6 3 Connecting TIO to the GFM using 5 10 Vdc output from RJ11 connector...
Страница 57: ...53 Figure 6 4 Connecting TIO to the GFM using 4 20 mA output from DB9 connector...
Страница 61: ...58 Figure 6 6 Connecting TIO to the GFC Using 0 5Vdc Input Output from DB15 Connector...
Страница 62: ...59 Figure 6 7 Connecting TIO to the GFC Using 4 20mA Input Output from DB15 Connector...
Страница 68: ...65 Figure 6 9 Connecting TIO to the Generic Flow Meter...
Страница 69: ...66 Figure 6 10 Connecting TIO to the Generic Flow Controller...
Страница 95: ...92 APPENDIX D Mechanical Drawings...
Страница 96: ...93 APPENDIX E Circuit Layout Diagrams Circuit Layout Top...
Страница 97: ...94 Circuit Layout Bottom Mirror...