31
e) Totalizer #2 Power On Delay (Numerical entry)
Sometimes it is convenient to start the Totalizer only after a specified
power-up delay interval. Most of the mass flow meters and controllers
require some warm-up time from the power-up event in order to stabilize
process variable output and to get accurate reading. “Totalizer #2 Power
On Delay” option allows setting a specified time interval which must
elapse from the device power-up event before the Totalizer will be
activated. Valid settings are in the range of 0 to 3600 seconds (default
value is 0, no delay).
f) Totalizer #2 Auto Reload (Tabular entry)
This option allows automatic reset/reload Totalizer #2 when it reaches
preset Action Volume value (when configured to count
UP
) or zero value
(when configured to count Down). This feature may be convenient for
batch processing when predefined volume of the fluid must be
repeatedly dispensed into the process. The following selections are
available: Enabled or Disabled. The default entry is Disabled. Totalizer #2
Auto Reload selections can be set with the
UP
and
DN
buttons and are
accepted by pressing the
ENT
button.
g) Totalizer #2 Auto Reset Delay (Numerical entry)
This option may be desirable when “Totalizer #2 Auto Reload” feature is
enabled. Valid settings are in the range of 0 to 3600 seconds (default
value is 0, no delay).
h) Reset Totalizer #2 (Numerical entry)
Totalizers #2 reading can be reset by selecting “Reset Totalizer #2” menu
option. A typical display with Totalizer #2 Reset screen is shown below.
Once the “YES” option is selected, Totalizer #2 will be reset and the
following conformation screen will appear.
Reset Totalizer #2:
NO
YES
DO YOU WANT
RESET TOTALIZER?
*******************
Totalizer Has
been reset!
Press any Key...
Содержание TIO
Страница 14: ...10 Figure 4 3 RS 485 Multidrop Half Duplex Two Wire System...
Страница 55: ...51 Figure 6 2 Connecting TIO to the GFM using 0 5 Vdc output from DB9 connector...
Страница 56: ...52 Figure 6 3 Connecting TIO to the GFM using 5 10 Vdc output from RJ11 connector...
Страница 57: ...53 Figure 6 4 Connecting TIO to the GFM using 4 20 mA output from DB9 connector...
Страница 61: ...58 Figure 6 6 Connecting TIO to the GFC Using 0 5Vdc Input Output from DB15 Connector...
Страница 62: ...59 Figure 6 7 Connecting TIO to the GFC Using 4 20mA Input Output from DB15 Connector...
Страница 68: ...65 Figure 6 9 Connecting TIO to the Generic Flow Meter...
Страница 69: ...66 Figure 6 10 Connecting TIO to the Generic Flow Controller...
Страница 95: ...92 APPENDIX D Mechanical Drawings...
Страница 96: ...93 APPENDIX E Circuit Layout Diagrams Circuit Layout Top...
Страница 97: ...94 Circuit Layout Bottom Mirror...