Troubleshooting
ID
C T Short text Full text
Corrective action
E172
E T
Time is expired
The expiry date of the QC
solution LOT is already past.
Register a LOT of QC solution
that is still valid.
E171
E T
Cannot export log.
Make sure that the USB flash
drive is connected properly
and that the system detects
it. If required, re-initialize the
USB port by tapping the 77
Elektronika logo in the top
right corner of the touchscreen
display.
E170
E T
Sample ID already exists,
please change it.
Verify and repeat the input or
use another Sample ID.
E169
E T
Registration Code is
already used.
Verify and repeat the input or
use another RegCode.
E168
E T
Registration Code is not
valid.
Verify and repeat the input or
use another RegCode.
E167
E T
Operator ID already exists,
please change it.
Enter another Operator ID.
E166
E T
Password check failed,
please try again.
Enter the valid password.
E165
E T
Password is too short,
please try again!
(minimum length is 3
characters)
Enter a new password that is at
least three (3) characters long.
E164
E T
Password does not match,
please try again.
Re-enter password.
E163
E T
Operator does not exist,
please try again.
The operator name is not on
the operator list. Enter another
Operator ID.
E162
E T
Operator disabled, please
try again.
The operator name has been
disabled. Enter another
Operator ID.
E161
E T
Sample ID required.
Please set it.
Enter Sample ID.
E160
E T
LOT Code required.
Please set it.
Enter LOT number from the test
strip package.
W169
W T
Cannot open serial port for
output!
Check serial port connection.
See point “8 The system does
not recognize one or more
external connectors (USB,
RS232, Ethernet, and so on).”
of the Troubleshooting chart.
W158
W T
Cannot open file for
output!
Check the output port and
presence of the output storage.