Testing
6 Testing
6.1 Measurement process
The analyzer can be operated in two different modes:
1. In normal mode, the system automatically waits for the strip to incubate for
one minute before it reads the first test pad. This is the default mode and the
throughput in this mode is approximately 50 strips per hour.
2. In fast mode, which can be selected at User Options, the test strip is meas-
ured directly after starting the test. In this case, it is up to the user to time the
incubation period outside the analyzer.
ǽ
For more detailed information regarding use and storage of test strips, please
refer to the strip’s instructions for use.
The test strip tray has to be correctly loaded into the reader. Prepare the test
strip, the urine sample and paper towel.
1
2
3
1. Dip the test strip into the urine sample, wetting all pads. Immediately remove
the strip from the urine.
2. Drag the edge of the strip against the side of the sample container.
3. Blot by touching the edge of the test strip to a paper towel to remove excessive
urine.
4
5
4. Place the test strip in the channel of the test strip tray with the test pads fac-
ing up.
5. The instrument will automatically detect an applied test strip. The measure-
ment cycle will be started. If the Autostart is deactivated the measurement
must be started using the
Start
button.
Do not use damaged strips.
Do not push or pull the test strip tray.