28
Instrument Settings
ǽ
The software does not check LOT code and expiry date entries. It is recom-
mended to double-check the entered codes.
ǽ
For more detailed information about the Options menu and User options,
please refer to the full User Manual (See chapter 1.4 How to use this manual).
10 Instrument Settings
The DocUReader 2 Pro device offers several settings to suit the specific workplace
requirements. System settings can be modified on the
Main » Options » Settings
screen.
ǽ
The list of available settings may vary according to the operator access level.
Fig. 17:
The Settings screen page 1 and 2
ǽ
For more detailed information about the instrument settings, please refer to
the full User Manual (See chapter 1.4 How to use this manual).
10.1 Output (Connectivity: Transfer/Export)
The DocUReader 2 Pro device can be connected
to other systems or storage devices by defining the
Output settings.
The system supports two protocols to transfer the
analysis results through an interface:
•
bidirectional (two-way) protocol based on the NC-
CLS LIS2-A2 standard protocol, the POTC1-A2 or
the HL7 protocol
•
unidirectional protocol, when the data are sent out
as a one-way data flow, either formatted
– as comma-separated values (CSV),
– or as UTF8 text.
The
Output type
text box (available after any of the
three output protocols is selected and
is tapped)
is used to define the communication port (available
options are based on the output protocol). Tap
to scroll through the list.
Fig. 18:
The Settings »
Output screen