Device installation
Do not disconnect the power ca-
ble while the device is in opera-
tion. Doing so can corrupt the
data and damage the system.
Make sure that there is no strip
on the test strip tray and that the
tray is clean before switching off
the device.
ǽ
If necessary (in case the system
freezes or the touchscreen fails),
switch off the device by pressing the
On/Off switch for at least five (5) seconds.
3.5.5 Calibrating the touchscreen
The touchscreen display is made of glass. Do not touch the screen if the
glass is cracked or shattered. Glass screens are sensitive to drops and
mechanical shocks.
The touchscreen of the device is correctly calibrated in the factory, but needs to
be recalibrated at least once a year. If the touchscreen is not responding or does
not respond accurately, complete the following steps to calibrate it:
1. Switch on or restart the device.
2. While the device is booting, wait for the progress bar at the bottom of the
screen to turn green. When it does, keep gently pressing down on the touch-
screen until the yellow calibration screen appears.
Do not use your finger to calibrate the touchscreen. Use a pointing de-
vice or a pen.
Do not use a pointing device that could damage the touchscreen such
as the tip of a pencil or the extended tip of a ballpoint pen.
3. Wait for the actual black screen calibration screen to appear. Tap the display
at the intersection of the crosshairs that appear in the corners and the center
of the screen with a pointing device. Try to tap the display as close to the
intersections as possible; this practice ensures the best possible alignment
between the touchscreen coordinates and the LCD screen behind it.
3.6 Software updates
ǽ
Only Administrator and higher access level operators can run a software up-
date.
The manufacturer is continuously upgrading the DocUReader 2 Pro user soft-
ware, adding new features and improving usability. From time to time the manu-
facturer will send a software update for the device. The following sections de-
scribe the software update procedure: