20
Testing
ǽ
The DocUReader 2 Pro will perform a sequence of checks (reference pad,
strip detection, position of slipped strip, dry strip, etc.) each time a test is run.
See “13.1 Strip checking events” for more information.
6
7
8
6. The strip position is checked before measurement.
7. A timer will count down the time remaining for analyzing the strip.
8. The analysis of the strip pads will begin.
To abort a measurement press the Back icon on the Analyzing screen
and press Stop/Drop on the Measurement screen.
Comments can be added during the countdown time.
After approximately 60 seconds the pad results will be displayed on the screen
and the test strip tray is automatically moved out of the analyzer.
The buttons remain inactive until the tray is fully moved out.
•
If Autostart is ON:
The result screen will be displayed until the test strip is
removed from the tray. Once the strip is removed, the display automatically
returns to the
Measurement
screen.
•
If Autostart is OFF:
The result screen will be displayed for approximately
5 seconds, while displaying a circle animation. Than the display will return to
the
Measurement
screen (if no error occurred during the readout). If the user
touches the display while the circle animation is displayed, the system will not
automatically return back.
9
10
11
9. Results Page 1/2
10. Results Page 2/2
11. Printed results