37
5. Insert 3M™ Steri-Gas™ EO Gas Cartridge, Load Chamber, and Close Door
Insert the 3M™ Steri-Gas™ EO Gas Cartridge into the cartridge bay located inside the GS Series sterilizer chamber (Figures 42). Push the green lever down over the Steri-Gas EO Gas Cartridge
to secure in place (Figure 43).
Sterilizer
22.2
˚
C
98.0
kPa
Open door
Cycle: 55 ˚C
Aeration (hh:mm): 12:00
Load ID: 120413-1
Cartridge lot number: 2017-02AB
Menu
Menu
Next
Next
Insert 3M
TM
Steri-Gas
TM
EO Gas Cartridge
Insert load
Close door
04/20/2015
05:26:08
6
4
5
2
3
4
1
7
Back
Back
Figure 41.
Step Five: Insert Steri-Gas EO Gas
Cartridge, Insert Load, Close Door
DANGER: To reduce the risks associated with exposure to ethylene oxide,
do not force the 3M™ Steri-Gas™ EO Gas Cartridges into the cartridge holder as excessive force could damage the cartridge and result in a cartridge leak.
CAUTION: To reduce patient risks associated with exposure to potentially non-sterile devices or inadequate sterilization procedures,
do not force the 3M™ Steri-Gas™ EO Gas Cartridges into the cartridge holder; excessive force could damage the cartridge and result in a cartridge leak.