23
Table 3 outlines the parameters in
Site Setup>Preferences
including description, default value, and minimum access level.
Parameter
Display Name
Parameter Description
Data Range (or
Format)
Default Value
Minimum
Access Level
to Edit
Enter PIN to
bypass scan
When enabled the system will require a Supervisor PIN to bypass scanning the
3M™ Steri-Gas™ EO Gas Cartridge during the cycle programming stage.
Disabled, Enabled
Disabled
Supervisor
Audible tone on
cycle failure
When enabled, system will produce an audible tone upon cycle failure. The audible
notification will sound at a pulsed rate of two (2) seconds on, one (1) second off, for total of
30 seconds.
Disabled, Enabled
Disabled
Supervisor
Audible tone on
cycle complete
When enabled, system will produce an audible tone upon cycle completion. The audible
notification will sound continuous for total of 20 seconds.
Disabled, Enabled
Disabled
Supervisor
Print on cycle
complete
When enabled, system will automatically print cycle reports on sterilizer printer.
Disabled, Enabled
Enabled
Supervisor
Cycle report
preference
Allows user to select default cycle report type for printing and saving.
Graph, Table, or Detailed
Graph
Supervisor
Sensor to plot
Allows user to select default sensors that will be plotted on graph of cycle reports.
Control, Monitor, Both
Control
Supervisor
Screen timeout
Allows user to select amount of time in minutes that must pass without a detected touch
before the system will dim the display.
[1:60] minutes
10 minutes
Supervisor
Idle mode
chamber
temperature
When enabled, the system will maintain user-entered target temperature while in Idle mode.
Off, GS [30 - 60] °C
Off
Supervisor
Pressure unit
Allows user to select preferred units for pressure on display and select reports.
mbar, kPa
mbar
Supervisor
User cycle count
Will count cycles performed between supervisor access level resets. Incremented at start of
a cycle.
[0 - 2147483647]
0
Supervisor- Reset
to zero only
Reset (User
cycle count)
When enabled with supervisor level access, will reset User cycle count indicator.
Disabled, Enabled
0
Supervisor- Reset
to zero only
Abated aeration
time
If the system is configured to use an Abator, this control will allow the user to enter how
many hours into aeration the system shall request the Abator to run. Abator start will remain
on for this amount of time after the start of aeration.
0 to 999:30 hours,
adjustable by
0.5 hour increments
0
Supervisor
Table 3. Site Setup Preferences Parameters
The
Site Setup>Cycle
tab provides the option to select the cycles that will appear on the Main Screen for the Operator, Supervisor, and Service. Figure 22 illustrates the options when selecting
the Operator Cycles for the Main Screen.
Sterilizer
22.2
˚
C
04/20/2015
12:37:05
98.0
kPa
Select cycle temperature:
6
5
4
3
2
1
7
55 °C
Return
Apply changes
Cycle access level
Touch check box to select cycles listed on start screen.
Touch row to change values.
Operator
Cycle
38 ˚C
55 ˚C
36.0 hrs
12.0 hrs
Minimum aeration
Setup
Preferences Cycle Setup Date-Time Network Site
Figure 22.
Setup Cycle Options for
Operator Cycles
Содержание Steri-Vac GS Series
Страница 1: ...3M Steri Vac Sterilizer Aerator GS Series Operator s Manual ...
Страница 2: ......
Страница 66: ......
Страница 67: ......