32
3M
™
Steri-Vac
™
Sterilizer / Aerator GS Series
– Operator’s Manual
12. Operating Instructions
12.1. Starting a Cycle
The Operator controls the 3M™ Steri-Vac™ Sterilizer/Aerator GS Series via a touch screen. A fingertip, stylus, ball tip pen or a computer mouse can be safely used to navigate the touch
screens. To use a computer mouse, simply plug the mouse into one of the two USB ports (reference Figures 6 and 7) and the mouse arrow will appear on the touch screen. There are seven (7)
programming steps to start a cycle.
DANGER: To reduce the risks associated with exposure to ethylene oxide,
inspect the display screen and cycle reports for errors and listen for audible notifications (if enabled). Always take action for errors as indicated in this manual.
CAUTION: To reduce patient risks associated with exposure to potentially non-sterile devices or inadequate sterilization procedures,
inspect the display screen and cycle reports for error codes and listen for audible notifications (if enabled). Always take action for error codes as indicated in this manual.
1. Select Cycle
The 3M™ Steri-Vac™ Sterilizer/Aerator GS Series has two validated preprogrammed cycle options: 38 °C (100.4°F) and 55 °C (131.0°F). Select the appropriate cycle. On a double door
sterilizer the Unload door must be closed to program a new cycle.
Sterilizer
22.2
˚
C
98.0
kPa
38 °C
Open door
Menu
Menu
Select cycle temperature:
55 °C
04/20/2015
05:19:51
6
5
4
3
2
1
7
Figure 33.
Step One: Select Cycle
Содержание Steri-Vac GS Series
Страница 1: ...3M Steri Vac Sterilizer Aerator GS Series Operator s Manual ...
Страница 2: ......
Страница 66: ......
Страница 67: ......