24
3M
™
Steri-Vac
™
Sterilizer / Aerator GS Series
– Operator’s Manual
The minimum default aeration time on the 3M™ Steri-Vac™ Sterilizer/Aerator GS Series can be changed for the 38 °C and 55 °C cycles. Press the minimum aeration time and an iWheel will
appear to adjust the minimum aeration default time. Figure 23 illustrates changing the minimum aeration default time.
Sterilizer
22.2
˚
C
04/20/2015
12:38:29
98.0
kPa
Select cycle temperature:
6
5
4
3
2
1
7
55 °C
Return
Apply changes
Cycle access level
Touch check box to select cycles listedon start screen.
Touch row to change values.
Operator
Cycle
38 ˚C
55 ˚C
2.0 hrs
2.0 hrs
Minimum aeration
Setup
Preferences Cycle Setup Date-Time Network Site
55 ˚C
Minimum aeration
10
11
12
00
hh:mm
:
30
13
14
00
:
12
00
:
12
OK
OK
Cancel
Cancel
Figure 23.
Changing the Minimum Aeration
Default Time
The
Site Setup>Setup
tab (Figure 24) provides the specific configuration and maintenance details for your site; this is established by authorized 3M service personnel during installation.
Sterilizer
22.2
˚
C
04/20/2015
12:39:27
98.0
kPa
Select cycle temperature:
6
5
4
3
2
1
7
55 °C
Return
Apply changes
Ventilation hood
High elevation
Abator always on
One door
Abator mode
Lock out immediate
Abator setting
1 to 1
Next maintenance
10-2015
Reset
Reset
Barcode override count
0
System cycle count
38
Last maintenance
04/20/2015
Maintenance interval
6 months
Setup
Preferences Cycle Setup Date-Time Network Site
Ventilation hood
High elevation
One door
Figure 24.
Site Setup>Setup Tab - Options