34
3M
™
Steri-Vac
™
Sterilizer / Aerator GS Series
– Operator’s Manual
Sterilizer
22.2
˚
C
98.0
kPa
Open door
Cycle: 55 ˚C
Aeration (hh:mm): 12:00
Menu
Menu
Next
Next
04/20/2015
05:21:59
6
4
5
2
3
4
1
7
Cancel
Cancel
Enter load ID:
1
4
7
.
abc
2
5
8
0
3
6
9
-
OK
OK
120413-1
Figure 36.
Load ID Entry by Keyboard
4. Scan 3M™ Steri-Gas™ EO Gas Cartridge Barcode
DANGER: To reduce the risks associated with exposure to ethylene oxide:
Only use 3M™ Steri-Gas™ EO Gas Cartridges as listed in this manual. Do not use other brands or types of sterilant cartridges.
Do not use damaged 3M™ Steri-Gas™ EO Gas Cartridges.
If an individual 3M™ Steri-Gas™ EO Gas Cartridge is ever dropped, the cartridge should be used immediately or disposed of as described in the cartridge disposal section.
WARNING: To reduce the risks associated with fire and explosion,
use care when handling 3M™ Steri-Gas™ EO Gas Cartridges as they contain 100% Ethylene Oxide (EO) which is an extremely flammable gas and liquid under pressure. Do not use near
flame, electrical sparks, or hot surfaces, or allow sources of ignition near the cartridges. Do not puncture cartridge outside the sterilization chamber. Do not incinerate cartridges. Exposure to
temperatures above 150ºF (65.5ºC) may cause cartridges to burst.
3M™ Steri-Gas™ EO Gas Cartridge
Catalog Number
3M™ Steri-Vac™ GS Series Models
Nominal Net Weight of EO
4-100
GS5 Series
EO Net wt. 100 g. (3.52 oz.)
8-170
GS8 Series
EO Net wt. 170 g. (5.99 oz.)
CAUTION: To reduce patient risks associated with exposure to potentially non-sterile devices or inadequate sterilization procedures,
only use 3M™ Steri-Gas™ EO Gas Cartridges as listed in this manual. Do not use other brands or types of sterilant cartridges.
Put on personal protective equipment (PPE) consisting of safety goggles, long sleeves, and butyl rubber gloves when handling full Ethylene Oxide (EO) Gas Cartridges. The barcode is a small
square, located on the top of the cartridge, in the area of the black stripe. To scan, place the barcode under the red light of the cartridge scanning bay on the front of the 3M™ Steri-Vac™
Sterilizer/Aerator GS Series (Figure 38). The barcode scanner will remain active for 10 seconds once scanning begins. After a successful scan, the 3M™ Steri-Gas™ EO Gas Cartridge lot
number will appear in the cycle parameters section of the GS Series sterilizer display screen.
Содержание Steri-Vac GS Series
Страница 1: ...3M Steri Vac Sterilizer Aerator GS Series Operator s Manual ...
Страница 2: ......
Страница 66: ......
Страница 67: ......