12.7. Printer Overview
The built-in printer provides easy-to-read information on each sterilization cycle. The cycle report is essential in analyzing the 3M™ Steri-Vac™ Sterilizer/Aerators GS Series performance and
can be retained to meet cycle verification policies.
To load printer paper, install the roll in the feed mechanism (Figure 56). Once the paper is fed into the mechanism, the printer will automatically advance the paper through the printer rollers.
Manually place the paper through the slot in the access door. To avoid damage to the print head, only use the paper intended for this printer.
Figure 56.
Loading Printer Paper
12.8. Unloading the 3M™ Steri-Vac™ Sterilizer/Aerator GS Series
DANGER: To reduce the risks associated with exposure to ethylene oxide:
Inspect the display screen and cycle reports for errors and listen for audible notifications (if enabled). Always take action for errors as indicated in this manual.
Always review the elapsed aeration time on 3M™ Steri-Vac™ Sterilizer/Aerator GS Series display prior to opening GS Series sterilizer door.
Never use force to access the inside of the sterilization chamber.
WARNING: To reduce the risks associated with fire and explosion:
Inspect the display screen and cycle reports for errors and listen for audible notifications (if enabled). Always take action for errors as indicated in this manual.
Do not attempt to access any internal mechanisms of the 3M™ Steri-Vac™ Sterilizer/Aerator GS Series. Operators should not service the sterilizer as there are no user serviceable parts.
CAUTION: To reduce the risk of injury,
follow good ergonomic practices. Loading baskets should not be overfilled requiring excessive force in pulling and pushing loaded baskets in and out of the sterilizer chamber. Reference facility
policies and procedures for appropriate ergonomic practices.
CAUTION: To reduce patient risks associated with exposure to potentially non-sterile devices or inadequate sterilization procedures,
inspect the display screen and cycle reports for errors and listen for audible notifications (if enabled). Always take action for errors as indicated in this manual.