3M ESPE Lava Chairside Oral Scanner User Guide - Beta
page 75 of 118
automatically flips its screen orientation. The direction of the wand and the direction of the
on-screen image thereby become consistent.
Pause Mode When Flipped
If you enter pause mode when the Maxillary Flip feature is active and you experience
difficulty recovering from pause mode, follow these instructions:
1.
The image will be aligned to the spot where you entered pause mode, so try to
recover from your original location.
2.
If that doesn’t work after a few seconds, return to an occlusal surface you have
previously scanned.
3.
If the paused image looks incompatible with the live image, press the Home Base
icon.
. That will “un-flip” the live image and make it consistent once again
with the motion of the wand.
Notes:
•
When Maxillary Flip is active, the Select button on the wand itself will also activate
the Home Base icon.
•
When Maxillary Flip is NOT active, there is little effect on the image when you press
the Home Base icon.
Inspection
You can use the wand’s camera to examine the field closely on the touch screen. It’s similar
to using an intra-oral video camera.
1.
Select the appropriate arch.
2.
Select the Inspection scan icon on the Scanning screen’s toolbar.
3.
Press the Start/Stop button to start the wand and inspect the area.
When you press the Start/Stop button again, the playback is displayed automatically. You can
review the video (see the illustration below) or you can press the Start/Stop button on the
wand to proceed directly to scanning.
Note
: The video captured is for review only and cannot
be saved.