3M ESPE Lava Chairside Oral Scanner User Guide - Beta
page 20 of 118
To clean the touch screen, bezel, wand body, and tether cord (and other cart surfaces as
needed):
1.
Apply the cleaning agent to a soft, non-abrasive cloth.
2.
Wring out the cloth until damp.
3.
Wipe down the surfaces with the damp cloth.
4.
Rinse the cloth in clean water.
5.
Wipe down the surfaces a second time.
6.
Dry the body of the wand and the tether cord with a clean, non-abrasive cloth. The
touch screen and bezel can air dry.
Do not spray or pour liquids on these surfaces.
Never submerge the scanning wand.
Do not use ammonia-based cleaners on these surfaces
.
Disinfection Procedures
While a patient is being scanned, the wand should be barriered. Approximately one inch of
the distal tip of the wand should be exposed during scanning so that the sheath does not
obscure the lens. Follow standard barrier procedures while using the scanner.
To prevent cross-infection, the tip of the wand must be disinfected before each use. Follow
these procedures:
1.
Unplug the wand from the dental workstation by pulling gently on the textured
portion of the connector from the back of the unit. (The unit itself can remain on.)