SEITE 2 - 148 x 210 mm - 44000118420/02 - SCHWARZ - 20-102 (kn)
2
4. Technical Data
Drive unit: Ball-bearing motor with
integrated self-resetting
overheat cut-off
Operating mode: Short-term operation,
1 min On/7 min Pause
Controller: Microprocessor-controlled,
high-precision mixing time
control by quartz timing
mechanism
Capsule holder: Self-locking
Capsules: Diameter max. 19 mm,
total length max. 58 mm,
length without tubes max. 38 mm,
weight max. 7 g
Housing: Housing and protective cover
made of high impact-resistant
plastics
Mixing speed: 4650 rpm for 7 g-capsules
Operating temperature: 18°C…40°C/64°F…104°F
Relative humidity: max. 80% at 31°C/88°F,
max. 50% at 40°C/104°F
Main voltage: 100 V, 120 V, 230 V, 50/60 Hz,
for preset voltage refer to the
rating plate
Power requirements: 330 VA
Fuses: 230 V: 2 A time-lag fuse
120 V: 3.2 A time-lag fuse
Dimensions: (W x H x D)
195 x 160 x 195 mm
Weight: 3750 g
Continous sound
pressure level: <70 dB(A)
Year of manufacture: See rating plate
5. Installation
Prior to the initial start-up, remove the transport
safety guard.
Connect the CapMix unit to the power supply and switch
on the unit with the power switch on the back of the
device. The display shows the mixing time of program 1.
The following basic settings are preset at the factory:
Program 1 (key “P1”): 5 sec mixing
Program 2 (key “P2”): 7 sec mixing
Program 3 (key “P3”): 9 sec mixing
Program 4 (key “P4”): 12 sec mixing
Fixed time (key “10 s”): 10 sec mixing
Fixed time (key “15 s”): 15 sec mixing
of the safety regulations may cause serious injury and
voids all warranties and guarantees.
14. Keep solvents, flammable liquids, and powerful heat
sources away from the unit as they may damage the
plastic housing..
15. Cleaning and disinfecting the unit, ensure that no
cleaning and disinfecting agent enters the device
since this may lead to electric short-circuiting or
hazardous malfunctions.
16. Pay attention to capsule leak tightness when mixing.
Capsules that are not leak tight contaminate the
device with mercury. Contaminated devices cannot be
repaired and must be disposed of as contaminated
waste according to local laws.
See the safety data sheet for the amalgam used for
first aid, measures to take in case of unintended
release, handling, storage, exposure limitation, and
personal protective clothing.
Please report a serious incident occurring in relation to
the device to 3M and the local competent authority (EU)
or local regulatory authority.
2. Product Description
The CapMix unit is a mixing device for dental filling
materials and cements in capsules. Desired mixing times
may be programmed into the mixing programs “P1” - “P4”.
Moreover, two pre-set mixing times, “10 s” and “15 s”, can
be selected.
These Instructions for Use should be kept for the duration
of product use. The product may only be used when the
product labeling is clearly readable. For details on all
additionally mentioned products please refer to the corre-
sponding Instructions for Use.
Intended Purpose
Intended purpose: mixing device for dental filling and
luting materials in capsules.
Intended users: educated dental professionals who have
theoretical and practical knowledge on usage of dental
products.
Patient target group: all patients requiring dental treat-
ment, unless the patient’s condition limits the use.
Contraindications
None
Potential Undesirable Side Effects and Complications
None known.
3. Fields of Application
Dental materials packaged in capsules with a max. capsule
weight of 7 g.
Capsules with a weight of more than 7 g must not be
used!
Содержание CapMix ESPE
Страница 2: ...SEITE U2 148 x 210 mm 44000118420 02 SCHWARZ 20 102 kn...
Страница 8: ...SEITE 6 148 x 210 mm 44000118420 02 SCHWARZ 20 102 kn 6...
Страница 26: ...SEITE 24 148 x 210 mm 44000118420 02 SCHWARZ 20 102 kn 24...
Страница 62: ...SEITE 60 148 x 210 mm 44000118420 02 SCHWARZ 20 102 kn 60...
Страница 68: ...SEITE 66 148 x 210 mm 44000118420 02 SCHWARZ 20 102 kn 66...
Страница 74: ...SEITE 72 148 x 210 mm 44000118420 02 SCHWARZ 20 102 kn 72...
Страница 80: ...SEITE 78 148 x 210 mm 44000118420 02 SCHWARZ 20 102 kn 78...
Страница 86: ...SEITE 84 148 x 210 mm 44000118420 02 SCHWARZ 20 102 kn 84...
Страница 93: ...SEITE 91 148 x 210 mm 44000118420 02 SCHWARZ 20 102 kn 91...
Страница 94: ...SEITE 92 148 x 210 mm 44000118420 02 SCHWARZ 20 102 kn 92...
Страница 95: ...SEITE U3 148 x 210 mm 44000118420 02 SCHWARZ 20 102 kn...