5.1 Setting the mixing time
Mixing times in the range of 1 sec to 39 sec may be
selected. By pressing the keys “+” or “-” the device
increases or decreases, respectively, the mixing time by
1 sec. Depressing and holding the “+” or “-” keys, the
device continues to count up or down, respectively.
5.2 Changing the basic settings
Set the mixing time as described above, then depress
one of the program buttons, “P1” - “P4”, until two short
signals are issued. This confirms that the desired settings
have been stored in the selected program.
6. Operation
6.1 Insertion of the capsule
Open and fold back the cover of the CapMix device. Press
down on the release button to release the capsule holder.
After the mixing time has elapsed, the latest time setting
entered is re-displayed so that there is no need to re-enter
any value if the same time is to be used.
The selected program can be interrupted at any time by
pressing the “Start/Stop”-button, upon which the program
returns to its original setting.
CAUTION: For safety reasons, the unit cannot be started
up unless the cover is closed. Attempting to operate the
unit with the cover being open causes the time display to
flash and an error message to be issued. Opening of the
protective cover during mixing is associated with a high
risk of injury. Hence, the mixing process is aborted. This
may render the capsule useless! Therefore, never open
the cover before the mixing process is completed and the
capsule holder has come to a complete stop!
6.3 Removal of the capsule
Open the lid of the CapMix unit.
Open the capsule holder by pressing the release button
down and towards the left, and remove the capsule.
7. Notes
Capsule materials of third-party manufacturers (including
amalgam capsules) may be mixed with the CapMix device
provided they are suitable for the device in terms of weight
and dimensions (please refer to the section, Technical
Data). The optimal mixing time for mixing with the CapMix
device should be inquired from the capsule manufacturer.
Pay attention to capsule leak tightness when mixing.
Capsules that are not leak tight contaminate the device
with mercury.
8. Operating Errors
Error Possible Remedy
cause
Time display is off, Power failure Check both the
even though the unit power cable and
is switched on main voltage
Motor fails to run Cover of the Close protective
when the unit is unit is open cover
switched on
Motor is Allow the unit
overheated to cool
NOTE: The CapMix device should be repaired by 3M or
an 3M-authorized repair facility only.
ENGLISH
en
Keep the release button depressed and push to the left to
open the capsule holder.
Place the activated capsule between the two cups of the
holding mechanism and release the release button. The
device clamps the capsule automatically. Insert capsules
with tips such that the tip points towards the left side.
Do not operate the device without a capsule inserted!
6.2 Mixing
Close the protective cover and press the “Start/Stop”-
button. This starts-up the mixing process.
The countdown of the mixing time is shown on the time
display.
Содержание CapMix ESPE
Страница 2: ...SEITE U2 148 x 210 mm 44000118420 02 SCHWARZ 20 102 kn...
Страница 8: ...SEITE 6 148 x 210 mm 44000118420 02 SCHWARZ 20 102 kn 6...
Страница 26: ...SEITE 24 148 x 210 mm 44000118420 02 SCHWARZ 20 102 kn 24...
Страница 62: ...SEITE 60 148 x 210 mm 44000118420 02 SCHWARZ 20 102 kn 60...
Страница 68: ...SEITE 66 148 x 210 mm 44000118420 02 SCHWARZ 20 102 kn 66...
Страница 74: ...SEITE 72 148 x 210 mm 44000118420 02 SCHWARZ 20 102 kn 72...
Страница 80: ...SEITE 78 148 x 210 mm 44000118420 02 SCHWARZ 20 102 kn 78...
Страница 86: ...SEITE 84 148 x 210 mm 44000118420 02 SCHWARZ 20 102 kn 84...
Страница 93: ...SEITE 91 148 x 210 mm 44000118420 02 SCHWARZ 20 102 kn 91...
Страница 94: ...SEITE 92 148 x 210 mm 44000118420 02 SCHWARZ 20 102 kn 92...
Страница 95: ...SEITE U3 148 x 210 mm 44000118420 02 SCHWARZ 20 102 kn...