TMA Grand Master 2.6 – User’s Guide
5 Using TMA Grand Master
Restore Medical Info Database
On the Restore Medical Info Database panel select the database that you want to restore,
then click
Restore
. The database then will be used as the basis for your project data.
Depths
Insertion depth (μm)
Depth of insertion into the recipient block can be set either by setting a negative or a positive
value, so that the tip of the puncher is positioned below or above the surface of the block
inserting the core at a required height. Value ranges from -500μm to +500μm.
Punching depth (μm)
Punching depth adjustment might be necessary to perform an accurate core removal
(depending on the location of the tissue within the block). Value ranges from -500μm to
+13000μm.
December 9, 2016 – Rev. 1
3DHISTECH Ltd.
74(92)