Model 106-H Ozone Monitor Manual Rev. D-2
23
Required calibration
parameters are large
(>
15% slope) when
calibrated using a
reliable ozone
instrument.
Flow path is contaminated.
Replace the absorption cell,
or return instrument to 2B
Technologies for cleaning or
replacement of the
absorption cell.
2B Technologies offers reasonably priced customer service for instrument repairs.
The calibration service includes cleaning of the entire flow path with methanol, testing
of all components for proper function and calibration against an IOA-traceable
standard. The best way to contact us for service is to log a customer service ticket at
www.twobtech.com/techsupport
. Normally, you will hear back from us by email within
a few hours. Or, call us at +1(303)273-0559.