Z-800 Instructions for Use.
7/69
P/N 800-600-01-08 Rev 08/10
GETTING STARTED
WARNINGS & CAUTIONS
The following Warnings and Cautions should be strictly followed to avoid harm to patients
and pump operators:
WARNING:
Zyno Medical assumes no responsibility for incidents that may occur
if its product is not used in accordance with its product labeling.
WARNING:
Only use Zyno Medical provided IV sets in the Z-800 pump.
WARNING:
The
Z-800 pump operation is strictly limited to trained operators
whose competency in safe Z-800 pump operation and safe IV therapy practices has
been tested and proven.
WARNING:
Make sure the pump is stable by fastening securely to an IV pole, or
resting on a flat surface. IV poles vary in quality and stability. Avoid fastening the
pump too high on the pole, and test for stability before using.
WARNING:
Prior to use, always verify the proper function of the display, audible
and visual alarms.
WARNING:
Verify there are no kinks in the tubing when loading into the pump.
WARNING:
Always read and follow the instructions that accompany the fluid
container and IV administration sets you are using. Carefully follow the
instructions in this document for loading, removing, and reloading the IV set.
Adjust the pumping section of the IV set every 24 hours, and replace the IV set at
least every 72 hours.
WARNING:
To prevent free flow, make sure to load the IV set in the pump before
connecting the IV set to patient.
WARNING:
To prevent free flow, make sure to disconnect the IV set from patient
before unloading the IV set from the Z-800 pump.
WARNING:
Disconnect IV set from the patient before purging air bubbles out of
IV tubing.
WARNING:
Some Z-800 IV sets configurations do NOT have set based free-flow
protection. Make sure to close roller clamp before removing administration set from
Z-800 pump.
WARNING:
Do not over-program VTBI. Program the actual amount of the fluid
in the IV bag.