Adjusting the Air-In-Line Alarm Threshold
The Z-800’s Air-In-Line sensor detects an air bubble in
the administration set tubing. Air-In-Line alarm
threshold can be adjusted in the
BIOMED OPTION
configuration menu. Alarm threshold can be adjusted to
air bubble size from a volume of 2 micro-liters to 280
micro-liters. The Z-800’s Air-In-Line Detector alarm
threshold can be adjusted by the user.
1.
From the configuration menu, Select
8. BIOMED
OPT
.
2.
In the BIOMED OPTION menu, Select
8.2. AIR
.
3.
In the AIR sub-menu, use
Up/Down Arrow
to adjust the air-in-line alarm threshold.
4.
Press
OK
soft key
to confirm the set value.
Press
QUIT
soft key
to back out of AIR sub
menu. Press
QUIT
soft key
again to exit the
BIOMED OPTION menu
WARNING:
The factory default settings
should be used unless qualified clinical personnel
determine that other customized settings are
appropriate and safe.