Z-800 Instructions for Use.
58/69
P/N 800-600-01-08 Rev 08/10
MAINTENANCE
BATTERY CARE AND MAINTENANCE (Continued)
Battery Care
The battery capacity should be checked at least once every
12 months.
If the Z-800 pump is to be stored at temperatures in excess
of 86 °F (30 °C) for one or more months, the battery should
be removed and placed in an environment between 50 –
86 °F (10-30 °C).
If the battery is to be stored for more than one year, they
should be charged at least once per year to prevent leakage
and deterioration in performance due to self-discharge.
When the battery is first being put into use, or has been out
of use for one or more months, it will not have full capacity
due to deactivation of reactants. Restore such battery to
original performance by repeating one or two cycles of
fully charging and fully discharging.
Some temporary reduction in capacity might become
apparent if the battery is repeatedly discharged less than
completely. One or two cycles of full discharge and full
charge can restore full performance.
The Z-800 Pump is shipped with a battery in a discharged
condition. Connect the power cord to an AC receptacle
and allow the battery to charge for 18 hours.
Whenever possible, leave the power cord connected to an
external AC power source while operating.
CAUTION:
Battery
replacement should be
performed by qualified
service personnel while the
instrument is not in use.
CAUTION:
All
pump configuration
settings need to be verified
and reset as needed after a
complete discharge of
battery.
CAUTION:
DO
NOT open, incinerate or
short circuit battery. Worn
out batteries must be
disposed properly,
according to local
regulations.