Z-800 Instructions for Use.
17/69
P/N 800-600-01-08 Rev 08/10
GETTING STARTED
CONTINUOUS MODE—PRIMARY INFUSION (Continued)
Change Primary VTBI
Use Up/Down Arrow keys to highlight Pri VTBI. Use
Data Entry keys to modify the corresponding digits of Pri
VTBI value.
Example
: Programming Pri VTBI = 250 ml
Use Up/Down Arrow keys to highlight Pri VTBI.
Press the 100’s Up data entry key
twice to increase
the 100’s digit of Pri VTBI to 200.
Press the 10’s Up data entry key
five times to
increase the 10’s digit of Pri VTBI to 50.
Start Primary Infusion
Verify displayed infusion parameter entries (Primary Rate,
Primary VTBI). If a clamp is engaged, remove the clamp.
Press
RUN/STOP
key to start the infusion.
Stopping and Resuming a Primary Infusion
Press
RUN/STOP
key during infusion. The infusion will be
paused. The
PAUSE
screen will be displayed.
From the
PAUSE
state, press
RUN/STOP
key to resume
infusion.
Titrate Flow Rate or VTBI During a Primary Infusion
From the
PAUSE
screen, press
PROG
soft key; the pump
will display primary infusion programming
screen. User is
then able to modify the primary RATE and primary VTBI
parameters.
RUN/STOP
key