Z-800 Instructions for Use.
26/69
P/N 800-600-01-08 Rev 08/10
GETTING STARTED
INFUSION COMPLETE
When the VTBI parameter of the current infusion
decreases to zero, the pump determines that the infusion is
completed and the pump automatically switches over to
KVO mode.
The Main display will present the ―INFUSION
COMPLETE – KVO‖ message and the KVO flow rate. An
audio warning tune will sound every 5 seconds until user
acknowledgement.
The Pump will continue infusing fluid into patient with a
―Keep Vein Open‖ rate of 5ml/h. The KVO rate of 5ml/hr
is also displayed in the Flow Rate LED.
Press
RUN/STOP
key to acknowledge the infusion
complete message. The pump will be paused and the KVO
infusion will be stopped.
WARNING: Some Z-800 IV sets configurations
do NOT have set based free-flow protection. Make sure
to close roller clamp before removing administration
set from Z-800 pump.