Z-800 Instructions for Use.
52/69
P/N 800-600-01-08 Rev 08/10
CONFIGURABLE SETTINGS (Continued)
Enable/Disable Drip Sensor Option
The Z-800 pump has an optional external drip sensor,
which may be used to detect a fluid container empty
event. The user may enable or disable the drip sensor in
the
BIOMED OPTION
configuration menu. When
enabled, the drip sensor will detect the absence of drips
in the drip chamber, continue to infuse 8mL, then alarm
and stop the pump.
1.
From the configuration menu, Select
8. BIOMED
OPT
.
2.
In the BIOMED OPTION menu, Select
8.7. DRIP
SENSOR
.
3.
In the DRIP SENSOR sub-menu, use
Up/Down
Arrow
to select ENABLE or DISABLE the
drip sensor.
4.
Press
SELECT
soft key
to confirm the set
value. Press
QUIT
soft key
to back out of
DRIP SENSOR sub menu. Press
QUIT
soft key
again to exit the BIOMED OPTION menu