Z-800 Instructions for Use.
19/69
P/N 800-600-01-08 Rev 08/10
GETTING STARTED
LOADING A SECONDARY ADMINISTRATION SET
1.
Program and start the primary infusion using a check-
valve primary administration set, as previously
described.
2.
Open a secondary administration set package, remove
the set and close the set roller clamp.
3.
Insert set spike into a prepared fluid container and hang
secondary container, following accepted clinical
procedures.
4.
Fill the drip chamber to 1/3 full.
5.
Open a secondary set and prime the set. Close the set
roller clamp.
6.
Attach a secondary set to the upper injection site on the
primary set.
7.
Ensure no air bubbles are present in the line.
8.
Hang the secondary fluid container at least 8 inches
above the primary solution container.
WARNING: Some Z-800 IV sets configurations do
NOT have set based free-flow protection. Make sure to
close roller clamp before removing administration set
from Z-800 pump.
WARNING: The bottom of the secondary
solution should be at least 19” above the top of the
pump.
WARNING: A minimum height differential
between primary and secondary solutions of eight inches
is essential for the safe operation of a primary/secondary
infusion.