Z-800 Instructions for Use.
65/69
P/N 800-600-01-08 Rev 08/10
APPENDIX
ALARM TESTING PROCEDURE (Continued)
Battery Test
1.
Connect the Z-800 pump to an approved AC power
outlet for at least 18 hours to allow the battery to fully
charge.
2.
Unplug the power cord from the AC power. Turn on
the device. Verify that the battery indicator
is ON
3.
Set a primary infusion with the following:
Pri RATE = 125 ml/h
Pri VTBI = 1000ml/h
4.
Start the infusion. Record the infusion starting time.
5.
Record the time when the Low Battery Warning is
presented.
6.
Verify the following:
Pump continues to operate during Low Battery
warning.
The Battery indicator
turned to red.
The warning audio tune and a visual message
―Low Battery, Plug in AC power‖ is presented
every 10 seconds.
7.
Record the time when the Battery Empty alarm is
presented. Record the current VINF, Pri RATE and
Pri VTBI parameters. Verify that the following:
Battery Empty alarm continues to sound until user
acknowledgement.
Infusion is stopped
8.
Allow the Battery Empty Alarm to continue to sound.
9.
Record the time the pump shut itself off.
10.
Connect the device to an approved AC power outlet.
11.
Turn on the device. Verify 12-14 below.