background image

9650-000840-01 Rev. B

SurePower

TM

 ll Battery Pack Guide

3

Using the SurePower II Battery Pack 

The 

SurePower II Battery Pack 

provides enhanced on-board processing that allows the battery 

to communicate its specific charging requirements to the 

SurePower

 Charger Station

 or 

ZOLL defibrillator/monitor unit. The battery’s on-board processing also allows it to provide the 
following information:

The battery’s run time or capacity.

The battery’s calibration status.

Battery Fault Indications.

Figure 1-1 ZOLL 

SurePower II Battery Pack 

Note:

ZOLL ships the 

SurePower II Battery Pack 

in a partially charged state and you must 

fully charge the battery pack before first use.

It is very important that you recharge a depleted 

SurePower II Battery Pack

 as soon as possible.

Caution

Do not store fully depleted battery packs or damage can result.

Caution

Do not store fully charged batteries for more than 90 days. Should you need to store a battery for 
90 days or longer, see the 

Extended Storage 

section of this guide for information on how to do 

so.

?

2:00+

3:00+

1:00+

®

3URE0OWER

4-

))

83ERIES

4-

0ROPAQ

š

--$

Summary of Contents for SurePower II

Page 1: ...SurePowerTM II Battery Pack Guide 2 00 3 00 1 00 T M T M 9650 000840 01 Rev B...

Page 2: ...er X Series and ZOLL are trademarks or registered trademarks of ZOLL Medical Corporation in the United States and or other countries All other trademarks are the property of their respective owners Co...

Page 3: ...use physical injury Damaged batteries may expose users to a safety hazard Should the SurePower II Battery Pack be dropped inspect and test the unit prior to further use following the procedures descri...

Page 4: ...battery from service as soon as practical and verify the correct operation of the charger defibrillator and battery Cautions Follow all recommended maintenance instructions If a problem occurs obtain...

Page 5: ...s calibration status Battery Fault Indications Figure 1 1 ZOLL SurePower II Battery Pack Note ZOLL ships the SurePower II Battery Pack in a partially charged state and you must fully charge the batter...

Page 6: ...tely 10 seconds to indicate the battery s estimated run time If the battery is fully charged the LED over each of the five runtime indicator icons illuminates As the battery discharges the number of i...

Page 7: ...r II Battery Pack Icon LED Indications This is the Recalibration icon This LED lights for approximately 10 seconds after you press and release the Display button if the battery needs to be calibrated...

Page 8: ...runtime indicator displays the battery runtime When you release the Display button the runtime indicator displays the battery capacity for 2 seconds then redisplays the battery runtime During the 2 se...

Page 9: ...ifies the battery fault there will be a brief delay before the LED blinks The fault codes number from 1 to 29 To indicate fault code 1 the Fault LED blinks once To indicate fault code 2 the LED blinks...

Page 10: ...Height in 17 8 Length x 9 3 Width x 3 Height cm Weight 1 4 lb 64 kg Voltage 11 1 V Capacity 6 6 Ah when charged to 12 6V Operating Temperature 32 F to 122 F 0 C to 50 C Optimal Charging Temperature 5...