9650-000840-01 Rev. B
SurePower
TM
ll Battery Pack Guide
5
Cleaning
Clean the
SurePower II Battery Pack
with a soft damp cloth and mild soap.
Do not immerse or set the
SurePower II Battery Pack
in fluids of any kind.
Do not
use alcohol
or ketones to clean the
SurePower II Battery Pack
.
Maintenance -- Testing and Inspection
Perform a battery check, which verifies that the battery is fully charged and fault free, as part of
any regularly scheduled unit test. Press the battery’s
Display
button to perform the battery
check.
As part of your defibrillator/monitor maintenance program, verify that the battery’s LEDs work
(press the
Display
button to test LEDs) -- if the LEDs do not work, remove the battery from
service. Also check the
SurePower II Battery Pack
casing for cracks, damaged contacts and
sockets, or other damage -- if the battery pack is cracked or damaged, remove it from service.
SurePower II Battery Pack
Icon
LED Indications
This is the
Recalibration
icon.
This LED lights for approximately 10 seconds (after you
press and release the
Display
button) if the battery needs to
be calibrated. If the
Recalibration
LED lights, the runtime
indicator will not display run time for that battery. For best
performance of the battery, you should recalibrate the
battery as soon as possible.
To manually recalibrate the SurePower II Battery Pack, you
can insert the battery into the SurePower Charger Station
(using the SurePower
TM
II Battery Adaptor) and perform a
ManualTest
(for more information, see the
ZOLL
SurePower Charger Station
Operator’s Guide
). After you
recalibrate the battery, the
Recalibration
LED will only flash
momentarily when you press the
Display
button.
This is the
Fault
Icon.
If the SurePower II Battery Pack detects a battery fault, this
LED lights for approximately 10 seconds after you press the
Display
button.
If the
Fault
LED lights, the runtime indicator will not display
run time for that battery.
See the following section for more information on battery
faults.