background image

I

NTRODUCTION

2

www.zoll.com

9650-000840-01 Rev. B

When in use on a defibrillator/monitor, if the battery pack indicates a fault condition or any 
operating state that can prevent the effective use of the defibrillator/monitor, replace that 
battery pack as soon as possible.

Do not place foreign objects into battery sockets.

Do not use the 

SurePower II Battery Pack 

in the presence of flammable agents (such as 

gasoline), oxygen-rich atmospheres, or flammable anesthetics. Using the device in the 
presence of flammable agents can cause an explosion. 

Do not immerse or set the 

SurePower II Battery Pack

 in liquid. Using the battery near or 

within puddles of water may present a shock hazard to the operator, patient, and nearby 
personnel.

Use only ZOLL chargers to charge ZOLL batteries.

If the display indicators on the 

SurePower II Battery Pack 

are not consistent with the 

indicators on the 

SurePower

TM

 Charger Station

, a fault condition may be present. Remove 

the battery from service as soon as practical, and verify the correct operation of the charger, 
defibrillator, and battery.

Cautions

Follow all recommended maintenance instructions. If a problem occurs, obtain service 
immediately. 

Do not use alcohol or ketones (MEK, acetone, etc.) to clean 

SurePower II Battery Packs

Do not sterilize or autoclave 

SurePower II Battery Packs

.

Avoid storage of fully depleted battery packs.

Do not install batteries into a defibrillator/monitor unit that is not plugged into live AC 
mains when storage may exceed 90 days. Battery damage may occur.

If you are disposing of a 

SurePower II Battery Pack

, contact your local authorities to 

determine the requirements for the recycling and disposal of electrical equipment and follow 
those requirements.

Notes

SurePower II Battery Packs

 may feel warm to the touch following a large number of shocks 

delivered at maximum energy or after recharging.

You can get additional information about the 

SurePower II Battery Pack

 and its use from the 

appropriate ZOLL defibrillator/monitor operator’s guide (for example, labeling information 
-- symbols used, Serial Number format, Date of Manufacture, and so on).

Summary of Contents for SurePower II

Page 1: ...SurePowerTM II Battery Pack Guide 2 00 3 00 1 00 T M T M 9650 000840 01 Rev B...

Page 2: ...er X Series and ZOLL are trademarks or registered trademarks of ZOLL Medical Corporation in the United States and or other countries All other trademarks are the property of their respective owners Co...

Page 3: ...use physical injury Damaged batteries may expose users to a safety hazard Should the SurePower II Battery Pack be dropped inspect and test the unit prior to further use following the procedures descri...

Page 4: ...battery from service as soon as practical and verify the correct operation of the charger defibrillator and battery Cautions Follow all recommended maintenance instructions If a problem occurs obtain...

Page 5: ...s calibration status Battery Fault Indications Figure 1 1 ZOLL SurePower II Battery Pack Note ZOLL ships the SurePower II Battery Pack in a partially charged state and you must fully charge the batter...

Page 6: ...tely 10 seconds to indicate the battery s estimated run time If the battery is fully charged the LED over each of the five runtime indicator icons illuminates As the battery discharges the number of i...

Page 7: ...r II Battery Pack Icon LED Indications This is the Recalibration icon This LED lights for approximately 10 seconds after you press and release the Display button if the battery needs to be calibrated...

Page 8: ...runtime indicator displays the battery runtime When you release the Display button the runtime indicator displays the battery capacity for 2 seconds then redisplays the battery runtime During the 2 se...

Page 9: ...ifies the battery fault there will be a brief delay before the LED blinks The fault codes number from 1 to 29 To indicate fault code 1 the Fault LED blinks once To indicate fault code 2 the LED blinks...

Page 10: ...Height in 17 8 Length x 9 3 Width x 3 Height cm Weight 1 4 lb 64 kg Voltage 11 1 V Capacity 6 6 Ah when charged to 12 6V Operating Temperature 32 F to 122 F 0 C to 50 C Optimal Charging Temperature 5...