Introduction

The 

SurePower

TM

 II Battery Pack

 is a rechargable, sealed lithium ion battery for use in 

compatible ZOLL

 

®

 defibrillator/monitors and chargers.

It is important that you read and understand the information in this manual. The proper use of 
the 

SurePower II Battery Pack 

is critical to the reliable operation of your ZOLL defibrillator/

monitor.

Conventions

This manual uses the following conventions:

Warning statements describe conditions or actions that can result in personal injury or 
death.

Caution

Caution statements describe conditions or actions that can result in damage to the 

SurePower II 

Battery Pack

.

Note:

Notes contain additional information on using and maintaining the 

SurePower II 

Battery Pack

.

The manual displays button and LED labels in 

bold

 face type.

Safety

The following sections provide important warnings, cautions, and notes regarding the ZOLL 

SurePower II Battery Pack

.

Warnings

Locate ZOLL defibrillator/monitor units where they cannot harm the patient or medical 
personnel if they fall -- accidental sliding or drop of the 

SurePower II Battery Pack

 may 

cause physical injury. Damaged batteries may expose users to a safety hazard. Should the 

SurePower II Battery Pack

 be dropped, inspect and test the unit prior to further use 

following the procedures described in the 

Maintenance

 section of this Guide. 

Avoid prolonged exposure of the 

SurePower II Battery Pack 

to direct sunlight to minimize 

heating and damage that may occur from exposure to ultraviolet light.

SurePower II Battery Packs

 have been tested against interference from radio frequency 

emissions typical of two-way radios used in emergency service/public safety activities. 
Refer to the appropriate ZOLL defibrillator/monitor operator’s guide to determine 
recommended operating distances from RF transmission equipment

Do not use 

SurePower II Battery Packs

 unless they have been visually inspected, tested, and 

recalibrated by appropriate personnel and you have been adequately trained in the proper 
use of these batteries. See the 

Maintenance 

section of this guide for more information.

Do not disassemble the 

SurePower II Battery Pack

.

Charge the 

SurePower II Battery Pack

 in a well-ventilated area.

You must perform periodic inspection, capacity testing, and functional testing of 
rechargable defibrillator batteries to ensure the availability of safe and reliable batteries. See 
the 

Maintenance 

section of this guide for more information.

Summary of Contents for SurePower II

Page 1: ...SurePowerTM II Battery Pack Guide 2 00 3 00 1 00 T M T M 9650 000840 01 Rev B...

Page 2: ...er X Series and ZOLL are trademarks or registered trademarks of ZOLL Medical Corporation in the United States and or other countries All other trademarks are the property of their respective owners Co...

Page 3: ...use physical injury Damaged batteries may expose users to a safety hazard Should the SurePower II Battery Pack be dropped inspect and test the unit prior to further use following the procedures descri...

Page 4: ...battery from service as soon as practical and verify the correct operation of the charger defibrillator and battery Cautions Follow all recommended maintenance instructions If a problem occurs obtain...

Page 5: ...s calibration status Battery Fault Indications Figure 1 1 ZOLL SurePower II Battery Pack Note ZOLL ships the SurePower II Battery Pack in a partially charged state and you must fully charge the batter...

Page 6: ...tely 10 seconds to indicate the battery s estimated run time If the battery is fully charged the LED over each of the five runtime indicator icons illuminates As the battery discharges the number of i...

Page 7: ...r II Battery Pack Icon LED Indications This is the Recalibration icon This LED lights for approximately 10 seconds after you press and release the Display button if the battery needs to be calibrated...

Page 8: ...runtime indicator displays the battery runtime When you release the Display button the runtime indicator displays the battery capacity for 2 seconds then redisplays the battery runtime During the 2 se...

Page 9: ...ifies the battery fault there will be a brief delay before the LED blinks The fault codes number from 1 to 29 To indicate fault code 1 the Fault LED blinks once To indicate fault code 2 the LED blinks...

Page 10: ...Height in 17 8 Length x 9 3 Width x 3 Height cm Weight 1 4 lb 64 kg Voltage 11 1 V Capacity 6 6 Ah when charged to 12 6V Operating Temperature 32 F to 122 F 0 C to 50 C Optimal Charging Temperature 5...

Reviews: