background image

U

SING

 

THE

 SurePower II Battery Pack

6

www.zoll.com

9650-000840-01 Rev. B

Extended Storage

Do not store the 

SurePower II Battery Pack

 for extended periods in a fully charged state. If you 

must store the battery pack for an extended period 

1. Charge the 

SurePower II Battery Pack

 until three runtime LEDs light (this takes 

approximately 1.5 hours when charging a fully depleted battery).

2.

Recharge the 

SurePower II Battery Pack

 until three runtime LEDs light every 4 months.

Checking Capacity Loss

To check battery capacity (the amount of charge the battery can hold), press and hold down the 

Display

 button for 

at least

 5 seconds. 

Press and hold down the 

Display 

button for at least 

5 seconds. 

While you hold down the 

Display

 button, the 

runtime indicator displays the battery runtime. 
When you release the 

Display

 button, the 

runtime indicator displays the battery capacity 
for 2 seconds, then redisplays the battery 
runtime.

During the 2 seconds after you release the 

Display 

button, 

the runtime indicator displays 

battery capacity and will light five LEDs to 
indicate that the capacity is equal or close to 
that of a new battery. 

After displaying battery capacity for 2 seconds, 
all LEDs blink and then display the battery’s 
runtime.

Over time, as you use a SurePower II Battery 
Pack
, the battery capacity decreases and this 
capacity check will light fewer LEDs. When the 
capacity check lights only a single LED, the 
battery is at or near 60% of its original capacity, 
and you should remove the battery from service.

If the 

Recalibration

 or 

Fault

 LED lights after 

you release the 

Display

 button, the runtime 

indicator will not display the capacity or run time 
for that battery.

Summary of Contents for SurePower II

Page 1: ...SurePowerTM II Battery Pack Guide 2 00 3 00 1 00 T M T M 9650 000840 01 Rev B...

Page 2: ...er X Series and ZOLL are trademarks or registered trademarks of ZOLL Medical Corporation in the United States and or other countries All other trademarks are the property of their respective owners Co...

Page 3: ...use physical injury Damaged batteries may expose users to a safety hazard Should the SurePower II Battery Pack be dropped inspect and test the unit prior to further use following the procedures descri...

Page 4: ...battery from service as soon as practical and verify the correct operation of the charger defibrillator and battery Cautions Follow all recommended maintenance instructions If a problem occurs obtain...

Page 5: ...s calibration status Battery Fault Indications Figure 1 1 ZOLL SurePower II Battery Pack Note ZOLL ships the SurePower II Battery Pack in a partially charged state and you must fully charge the batter...

Page 6: ...tely 10 seconds to indicate the battery s estimated run time If the battery is fully charged the LED over each of the five runtime indicator icons illuminates As the battery discharges the number of i...

Page 7: ...r II Battery Pack Icon LED Indications This is the Recalibration icon This LED lights for approximately 10 seconds after you press and release the Display button if the battery needs to be calibrated...

Page 8: ...runtime indicator displays the battery runtime When you release the Display button the runtime indicator displays the battery capacity for 2 seconds then redisplays the battery runtime During the 2 se...

Page 9: ...ifies the battery fault there will be a brief delay before the LED blinks The fault codes number from 1 to 29 To indicate fault code 1 the Fault LED blinks once To indicate fault code 2 the LED blinks...

Page 10: ...Height in 17 8 Length x 9 3 Width x 3 Height cm Weight 1 4 lb 64 kg Voltage 11 1 V Capacity 6 6 Ah when charged to 12 6V Operating Temperature 32 F to 122 F 0 C to 50 C Optimal Charging Temperature 5...

Reviews: