Understanding Battery Faults

The 

SurePower II Battery Pack 

can detect one or more simultaneously occurring faults that 

render the battery unfit for use.

To determine the faults that affect a battery, follow this procedure when the fault light is on:

The following table lists battery temperature faults and actions you can take to clear these fault 
conditions. All other fault codes require you to remove the battery from service. For more 
information on these fault codes, see the appropriate ZOLL defibrillator/monitor service 
manual. 

Press and hold down the 

Display 

button for at least 

15 seconds.

If the battery does not detect any battery faults, all of the 
LEDs will light for 10 seconds.

The 

Fault 

LED will blink to indicate the code that identifies 

the battery fault (there will be a brief delay before the LED 
blinks).

The fault codes number from 1 to 29. To indicate fault code 
1, the 

Fault 

LED blinks once. To indicate fault code 2, the 

LED blinks twice, and so on.

If the battery detects multiple battery faults, it displays the 
first fault code, pauses briefly, and then displays the next 
fault code, and so on.

Fault ID

Fault Type

Recommended Action

13

Battery Over Temperature 
During Charge.

The battery is at or above its 
maximum temperature for 
charging (140°F/60°C).

Move the battery to a cooler room or out of 
direct sunlight, if possible, and allow the 
battery to cool before continuing to charge the 
battery. 

If the ambient temperature was low when the 
battery reached or exceeded its maximum 
temperature, do not use this battery. 

14

Battery Under Temperature 
during Charge.

The battery is at or below the 
minimum temperature for 
charging (32°F/0°C).

Move the battery to a warmer room (above 
32°F / 0°C) and allow the battery to warm up 
before charging again.

Summary of Contents for SurePower II

Page 1: ...SurePowerTM II Battery Pack Guide 2 00 3 00 1 00 T M T M 9650 000840 01 Rev B...

Page 2: ...er X Series and ZOLL are trademarks or registered trademarks of ZOLL Medical Corporation in the United States and or other countries All other trademarks are the property of their respective owners Co...

Page 3: ...use physical injury Damaged batteries may expose users to a safety hazard Should the SurePower II Battery Pack be dropped inspect and test the unit prior to further use following the procedures descri...

Page 4: ...battery from service as soon as practical and verify the correct operation of the charger defibrillator and battery Cautions Follow all recommended maintenance instructions If a problem occurs obtain...

Page 5: ...s calibration status Battery Fault Indications Figure 1 1 ZOLL SurePower II Battery Pack Note ZOLL ships the SurePower II Battery Pack in a partially charged state and you must fully charge the batter...

Page 6: ...tely 10 seconds to indicate the battery s estimated run time If the battery is fully charged the LED over each of the five runtime indicator icons illuminates As the battery discharges the number of i...

Page 7: ...r II Battery Pack Icon LED Indications This is the Recalibration icon This LED lights for approximately 10 seconds after you press and release the Display button if the battery needs to be calibrated...

Page 8: ...runtime indicator displays the battery runtime When you release the Display button the runtime indicator displays the battery capacity for 2 seconds then redisplays the battery runtime During the 2 se...

Page 9: ...ifies the battery fault there will be a brief delay before the LED blinks The fault codes number from 1 to 29 To indicate fault code 1 the Fault LED blinks once To indicate fault code 2 the LED blinks...

Page 10: ...Height in 17 8 Length x 9 3 Width x 3 Height cm Weight 1 4 lb 64 kg Voltage 11 1 V Capacity 6 6 Ah when charged to 12 6V Operating Temperature 32 F to 122 F 0 C to 50 C Optimal Charging Temperature 5...

Reviews: